Processing Record for this file: (Make notes here unique for this file) Loading from 2015-05-01 - 2015-06-01 Data taken from DMF, untarred, unzipped and renamed back to orignal file names It was noted that BOTH TEI 42i's lost their internal Mass Flow Meters. This was not consequential as the inlet has much better Mass Flow Controllers. Ambient data appear valid. Diurnal cycles of all three species are readily apparent. Multiple large (>100 ppbv) excursions noted. These correlate very well with CO excursions. They are very fast and so the NO2 values (which depend on a difference measurement) are not valid. It is nearly impossible to discriminate these periods. It is concluded that these periods MAY be due to local generator emissions and the associated variability with instrument/pump restarts. NOx photolysis cell appears spot on with conversion efficiency. Based on calibrations, the NO detector efficiency used in calculating concentrations was changed from 1.070 to 1.182. 1.070/((.891+.919)/2) Based on calibrations, the NOx detector efficiency used in calculating concentrations was changed from 1.087 to 1.182. 1.087/((.907+.933)/2) Based on calibrations, the NOy detector efficiency used in calculating concentrations was changed from 1.506 to 2.34. 1.506/((.6392+.648)/2)